Time Interval Analyzer 7 Analog Modules
System Reference, January 2001
229
• Propagation Delay Measurement
This mode is used to measure the time difference between
edges of two input channels:
Figure104
TrigDelayCnt
Parameters for Propagation Delay Measure-
ment
For propagation delay measurement, the TIA counts edges
determined by the measurement mode for start channel
and stop channel, respectively.
In the above example, as
TrigDelayCntStart(1)
and
TrigDelayCntStop(2)
are specified, the TIA measures
the propagation delay from the first rising edge of the start
channel signal to the second rising edge of the stop
channel signal after the trigger.
Trigger Mode and Number of Measurements
The TIA starts the interval measurements when the speci-
fied edge (rising or falling edge) of the trigger is detected.
The TIA has two trigger modes the auto trigger mode and
the external trigger mode.
PD++ (PD +, PD+ ,PD )
Trigger
Start Channel
Signal
1 2TrigDelayCntStart(1)
TrigDelayCntSto
The TIA measures this interval.
Stop Channel
Signal
1 2
3