4 Test Head Filling and DUT Board Considerations Test System Configuration
82 System Reference, January 2001
The fill order of the cages is marked above the brackets at
the top of the group scheme in Figure 40.
NOTE: This is the fill order to follow. The fill order
mentioned in the drawing E6980-96550-1S3 is INVALID.
The slot number of the slots, where an analog module has
to go within the cage, is mentioned to the left of the half
groups.
A TIA test head card always comes with a GP TIA and
either with a second GP TIA or, optionally, with a HP-TIA.
The first card containing a GP TIA is placed in slot 1 of
group 8 and the second one is placed in the first slot of
group 6. In each of the two slots, you can chose one of the
two optional TIAs on top ( marked “slot b” in the table in
Table 6). This way you can obtain a maximum number
(MAX) of 4 GP TIAs. Otherwise, you can have up to 2 HP
TIAs but in this case only 2 GP TIAs.
In the case of the sampling instruments, there are two
samplers in a slot as the samplersare dual and thus two
count numbers each are assigned in Figure 40. Therefore,
the maximum number of single samplers is additionally
mentioned in brackets in the table in Table 6.
NOTE that group 6 supports MEASurement only to
increase the number of digitizers without compromising
the number of digital channels. The Agilent source chan-
nels (except UHFAWG) have buffered outputs so that fan
out to multisite is made simple. Therefore, more digitizers
are desireable for multisite and allowing MEASurement
to be supported in an entire group makes this possible.
The blank slots in Figure 40 are kept for future enhance-
ments.