Synchronization 7 Analog Modules
System Reference, January 2001
243
Table 44 Trigger-to-Signal Delay, Accuracy, and Uncertainty
Where, “1 master clock period” means one period of the
master clock used for the analog module.
Accuracy” in the above table means a fixed delay that
could not be compensated by the analog module’s timing
calibration. Further, it has repeatability; if a test is done
under certain conditions of the analog module, the analog
module always starts “Trigger-to-Signal Delay
± the fixed
delay” after receiving the trigger signal when the “Uncer-
tainty” is zero.
Uncertainty” means the dispersion of the start timing of
the analog module. It has no repeatability; the start timing
of the analog module may be different even if the test is
done under the same conditions. The analog module starts
in the uncertainty range around the “Trigger-to-Signal
Delay
± the fixed delay” point after receiving the trigger
signal. See the following figure.
Analog Module
Trigger-to-
Signal Delay Accuracy Uncertainty
High Resolution AWG 1000 ns ±250 ns ±1 master clock period
High Speed AWG 200 ns ±1 ns 0 to ±1 master clock pe-
riod
Ultra High Speed AWG 110 ns ±(450 ps + OTA) ± 1 sampling clock
High Resolution Digi-
tizer
0 ns ±500 ns ±1 master clock period
High Speed Digitizer 50 ns ±1 master clock
period
0 to ±1 master clock pe-
riod
Dual High Speed Sam-
pler
70 ns ±1 ns 0 to ±1 master clock pe-
riod
Vista de página 242